Time series of the secondary electron yield were recorded using a
personal computer controlling a Keithley 6517A electrometer through
an IEEE 488 bus. For a complete measurement cycle the bias voltage
and the voltage
on one of the deflection
plates needed to be changed by software. A special interface was built
for this task (Figure 3.2).
To switch the beam on and off the voltage applied to
one of the deflection plates was switched between two values provided
by two power supplies. These deflection plates are located at the
exit of the electrostatic analyzer. To avoid problems arising from
the electrical connection between the deflection system and the controlling
PC the switch was realized by relays providing galvanic insulation
between the two systems. For the sample bias voltage
it was impossible to use relays in the same way as for the deflection
voltage
because of the very small current to be measured.
Even very small leakage currents through some mounting structure towards
ground had to be avoided. This switch was realized using a battery
and some electronics mechanically attached only to the vacuum interface
pin. A battery powered relay controlled through a light dependent
resistor (LDR) was used to switch between
=0V and
=18V.
The only connection to ground is through the ampere meter. A light
emitting diode (LED) illuminates the LDR to switch the relay. The
LED has no mechanical contact at all with the switching electronics
providing a maximum of insulation. Power to the switching electronics
was provided by a two batteries, one for the bias voltage itself and
the other supplied the driver and the relay. The LED and the relay
for
were controlled via the parallel port of the PC.
Figure 3.3 depicts a typical measurement cycle. A total of 16 frames of 100 measurements each are recorded with alternating settings for beam deflection and sample bias voltage. Appropriate time delays between the frames ensure that charge displacements on insulated parts of the chamber can take place without any negative influence on the current measurement. The observed time constants suggested a replacement circuit for the current measurements as shown in Figure 3.4.
![]()
|
At the start of each of the 16 frames a delay of three seconds allows the whole system to stabilize after a switch of the primary beam. No data is taken during this time. The time constant
A value for was estimated by
![]() |
(3.12) |
where is the change in the measured current when
is switched while the primary beam is not changed (see
Figure 3.3). This yields a value of
.
Another problem for time resolved measurement is the non shielded part of the circuit shown as shaded area A in Figure 3.4. Induction might cause additional currents in this loop introducing additional noise. By keeping the area small and by avoiding movements in proximity of the circuit during measurements the induced current can be kept low. Useful measurements may only be taken during periods of quiet laboratory environment. Some unexplained temporarily current fluctuations may be due to the nearby air conditioning system for the laboratory below or due to traffic in the nearby railway station.
March 2001 - Martin Wieser, Physikalisches Institut, University of Berne, Switzerland