To characterize the properties of the lens measurements using ions
were performed. The conversion surface was replaced by a highly transparent
grid. Because of the positive ions generated in the CASYMS ion source
the lens voltages were reversed and the grid set to ground potential.
O
ions made out of CO
were used as primary particles.
O
ions from 10eV to 450eV were used for the measurements.
Using the x-table the entry slit of the CODIF was moved across the
exit aperture of the extraction lens to investigate the primary energy
focusing properties of the lens. Furthermore, the overall voltage
over the lens was varied from 5kV to 15kV in order to find a combination
of parameters with a high signal to noise ratio. Keeping the primary
beam energy constant, the location of maximum intensity was shifted
towards the low energy edge of the lens exit aperture when increasing
the overall lens voltage. Two maxima were observed when varying the
lens position (x-translation). The splitting into two maxima was probably
caused by beam deflections at the grid. With 0.5mm spacing between
the wires of the grid the path of a low energy ion (E < 100eV) might
have been disturbed easily when getting close to the grid, resulting
in particles leaving the grid on the inside at angles not representative
of converted neutrals.
March 2001 - Martin Wieser, Physikalisches Institut, University of Berne, Switzerland